Safety checks could reduce radiotherapy errors
A combination of several well-known safety procedures could greatly reduce patient-harming errors in the use of radiation to treat cancer, according to a new study led by Johns Hopkins researchers.
Radiation oncologists use more than a dozen quality assurance checks to prevent radiotherapy errors, but until now, the Johns Hopkins researchers say, no one has systematically evaluated their effectiveness. Working with researchers at Washington University in St. Louis, the Johns Hopkins team gathered data on about 4,000 “near miss” events that occurred during 2008 and 2010 at the two institutions. They then narrowed the data set to 290 events in which errors occurred that—if they had not been caught in time—could have allowed serious harm to patients. For each commonly used quality assurance, or QA, check, they determined the percentage of the potential patient-harming incidents that could have been prevented.
The group’s key finding was that a combination of approximately six common QA measures would have prevented more than 90 percent of the potential incidents.
“While clinicians in this field may be familiar with these quality assurance procedures, they may not have appreciated how effective they are in combination,” said Eric Ford, assistant professor of radiation oncology and molecular radiation sciences at Johns Hopkins, who presented the group’s findings Aug. 3 at the joint American Association of Physicists in Medicine and Canadian Organization of Medical Physicists annual meeting, held July 31 to Aug. 4 in Vancouver, Canada.
At a separate symposium at the meeting, also on Aug. 3, Ford and his colleagues made related recommendations for the standardization of radiotherapy accident investigation procedures.
Ionizing radiation, such as gamma radiation or proton beam radiation, has long been a staple in cancer treatment because it can efficiently create cell-killing DNA breaks within tumors. The goal is to use it in ways that maximize the dose delivered to a tumor while keeping healthy tissue around the tumor as protected as possible by sharply focusing the radiation treatment area.
Unfortunately, the multistep complexity of radiation therapy and the numerous precision measurements its use entails can sometimes lead to mistakes, with patients getting too little radiation where it’s needed, or too much where it isn’t.
One QA check, a piece of hardware called an Electronic Portal Imaging Device, or EPID, is built into many radiotherapy-delivery machines and can provide a real-time X-ray-like image of the radiation coming through a patient. But Ford says that less than 1 percent of radiotherapy clinics use EPID because the software and training needed to operate it are mostly absent.
However, Ford says, the team’s research showed that another key to safety turned out to be a humble checklist of relatively low-tech measures, “assuming it’s used consistently correctly, which it often isn’t.” The checklist includes before-treatment reviews of patient charts by both physicians and radiation physicists, who calculate the right dose of radiation.
Use of film-based radiation-dose measurements as an alternative to EPID and a mandatory “timeout” by the radiation therapist before radiation is turned on to double-check that the written treatment plan and doses match what’s on the radiation delivery machines were also on the list of the most effective QA procedures.
A common QA measure known as pretreatment IMRT (for intensity modulated radiation therapy), in which clinical staff do a “test run” of the radiotherapy device at its programmed strength with no patient present, ranked very low on the list because it would have prevented almost none of the potential incidents studied. “This is important to know, because pre-treatment IMRT often consumes a lot of staff time,” Ford said.
Ford and his Johns Hopkins colleague Stephanie Terezakis, a pediatric radiation oncologist and a contributor to the QA evaluation study, are members of the American Association of Physicists in Medicine Working Group on the Prevention of Errors. In an Aug. 3 symposium at the Vancouver meeting, the group made recommendations for a national radiotherapy incident reporting system. The group is developing a way to have treatment errors and near misses reported and sent to a central group for evaluation and dissemination to clinics, Ford says. “It could work in ways similar to how air and train accidents are reported to the National Transportation Safety Board,” he noted.
Other experts who contributed to the QA-check effectiveness study are Kendra Harris and Annette Souranis, both of Johns Hopkins; and Sasa Mutic, of Washington University in St. Louis.
The study was funded with a pilot research grant from Elekta Inc.